- 5 µH Line Impedance Stabilisation Network for EMC conducted emission testing
- Ensures stable and standardized impedance conditions during measurements
- Designed for accurate EMI compliance testing of electronic devices
- Wide frequency range suitable for international EMC standards
- High reliability design with precision internal components
- Solid metal enclosure for effective shielding and noise suppression
- Easy connection to EMI receivers and spectrum analyzers
- Ideal for EMC laboratories, R&D, and industrial test environments
Line Impedance Stabilisation Network
The TBL05100-1-B is a universal 5µH LISN that meets the requirements of several standards. The TBL05100 is factory jumpered to 50Ω // 5µH + 1Ω, fulfilling the CISPR 16-1-2 and CISPR 25 impedance, phase an isolation specification. By setting internal jumper, it can be converted into 50Ω // 5µH variants that support standards such as CISPR 25, MIL-STD-461G, ISO11452-4 and ISO 7637-2. By adding an external 10F capacitor, the impedance will fulfil DO-160 specifications. The LISN is inserted into the EUT’s (Equipment Under Test) supply line. Conducted noise at the EUT’s supply terminals can be monitored at the BNC connector with a spectrum analyzer or a measurement receiver. A 5µH inductor decouples the source (supply) terminal from the EUT terminal. The use of two TBL05100-1-B in conjunction with the Tekbox LISN Mate allows for the measurement of common mode and differential mode noise separately. The unit comes equipped with a BNC-male to N-male RG232 cable, mating connectors and ground brackets to attach it to a ground plane.
Features
Key capabilities and advantages
Specification
Technical specifications and parameters
| Topology | Single path, configurable, 50Ω // 5µH + 1Ω, 50Ω // 5µH; internal capacitor can be disconnected for ISO 7637-2 and DO-160 setups; external 10µF capacitor required for DO-160; factory setup is 50Ω // 5µH + 1Ω; other configurations require internal high-current jumper setting |
| Supported standards | CISPR 16-1-2, CISPR 25, MIL-STD-461G, DO-160, ISO 11452-4, ISO 7637-2 |
| Characterized frequency range | 10 kHz – 400 MHz |
| DC resistance | < 5 mΩ (Source+ to EUT+), < 5 mΩ (Source− to EUT−) |
| Maximum current | 100 A (see heat-up characteristics plot, Chapter 8) |
| Maximum operating voltage range | 0 – 500 V AC @ 50/60 Hz, 250 V AC @ 400 Hz, 1000 V DC |
| High current plug / screw terminals | Male – Phoenix Contact 1762741; Female – Phoenix Contact |
| Accessories supplied | 2 pcs female terminal blocks |
| Jumper | Harwin D3087-98 |
| Dimensions | 300 mm × 150 mm × 150 mm |
| Weight | 2.7 kg |